SALTS IN CRUDE OIL (ELECTRONIC METHOD)
Description :
Fully automatic, the NSB TECH offers a precise test of salt concentration in crude oil. This test measures conductivity in the crude oil due to the presence of common chlorides, such as sodium, calcium, and magnesium by electronical methods.
Knowledge of the chloride concentration gives information on the efficiency of the desalination process and provides a solid basis for assessing the need for crude oil desalting. The corrosion rates in refining units depend on the chloride concentration.
Characteristics
Fully automatic model
Plug and play
Less than 30 sec for measurement
Digital auto ranging meter to read the voltage and current
Auto voltage adjustment between 0 and 250 VAC
Accuracy for the calibration (± 0.5%)
Salt concentration values displayed automatically in PTB
g/m3 or mg/kg
Auto-selection of ASTM
High current resolution (0.01 mA)
Security
High level of safety
Safety for unintended touching of electrodes
Over current safety
Hall effect sensor to cut the power with the lid opening
Alarm during measurements
Light indicator
Features
The NSB TECH is fully automatic. Easy to use, the test is done alone with 5 preset voltages. The correlation is established through 3 pre calculated record tables. The table is selected depending on the sample temperature. This unit offers a digital readout of both current and voltage for easy
checking. The device gives the best results recorded. The data is printed or exported to USB memory stick. Customizable, it is easy to set up the test data or create a custom table. The 50 last tests are archived.
To maximize user safety, a Hall effect sensor checks contact between the cover and the tank. A green indicator light warns of good positioning and a red light indicates when the unit is on. An audible alarm warns during the test of the 5 measurements.
Accessories
Suitcase
Calibration kit
Delivered with
1 Cover with electrode
3 Beakers with rings
Technical data
Power supply: AC 115V or 230V, 50/60Hz – 1A
Dimensions: (W) 250 x (D) 380 x (H) 290 mm
Weight: 7kg
Recommendations
- Chemical products requirements for analyse : neutral oil (refine chloride-free oil, app. 20mm²/sec viscosity at 40°C and free of additive)
- butanol
- absolute methylic alcool (anhydrous)
- water distillate, xylene (reageant grade, minimum purity)