EXISTENT GUM IN FUELS BY JET EVAPORATION
Description :
The device determines existent gum in fuels by steam-jet and air-jet evaporation. There is exclusive triple digital controllers for : air temperature (155 +/- 0,5 °C), steam temperature (232 +/-2 °C), and both air & steam flows (1000 mL/s +/- 15 %).
Characteristics
Temperature range up to 260°C
Switch air / steam
Small footprint
5 positions
Electronic temperature control
Easy installation of the 5 injectors
Integrated superheater for steam method
Security
Powercut in case of overheat
Accessories
Air compressor: 2 stages suitable to deliver correct flow of air, 300 l capacity. For use on
AC 400 V III, neutral (without plug), 7 A, (1670x600x760mm) – 125 kg
Steam generator: (400V – 3 phases) for use with ref. 941320 apparatus, equipped to maintain vapour output at requested quantity, automatic water feeding with suppressor, automatic water level control, two safety devices (pressure switch and safety valve), manometer, liquid level and supplied with flexible connecting pipe. Supply : 400V, III,
neutral (without plug), 50-60 Hz – 0,55 kW – (550x440x980mm) 72 kg.
Analytical balance: 120 grams (0,1 mg)
Delivered with
• 1 Thermometer 3C * without certificate
• 5 Beakers
• 5 conical outlets
• 1 gripp
Technical data
Power supply: 230 V, 50/60 Hz – 16A –
Dimensions: (W)580x (D) 350x (H)440mm
(W) 580x (D) 350x (H) 630 mm with grip
Weight: 40 kg
Recommendations
Service pressure (bar): 2,15 bars max compressed air
* certificate on request