Cleanliness

NMC 460

Micro carbon conradson residue tester - automatic

Ref. : 61100
  • Holding temperature : 450°C to 550°C
  • Type K thermocouples (oven and safety)
  • Routine analysis for carbon residues: from 0.10% to 30%
  • Heating rate of 12.5°C/min (±2.5°C/min)
  • Up to 12 samples can be analyzed simultaneously
  • Automatic calculation of carbon residue
  • Oven cooling by internal turbine
  • Nitrogen flow from 150 to 600ml/min with 2 preset flows and mass flow meter control
  • 1 fixed program (compliant with standard)
  • 12, 7 and 6-hole carousels
  • Glass vials : 2ml / 4ml / 16ml
  • Pressure gauge on the front panel
  • Purge phase adjustable up to 50 minutes
  • Forced ventilation of the electronic compartment
  • Gas fitting for Ø 4mm tubing, 9 bar max
  • Calibration in real conditions (or by simulator) ± 9.9°C
Supplied with :
  • 12 vials of 2mL
  • 6 vials of 16mL
  • 3 holders : large, small & mixed
  • Lid gripping hook
  • Cleaning cable
  • Insulating lid
  • 1 pair of high-temperature gloves
Options and accessories
  • Small vials, 2mL
  • Medium vials, 4mL
  • Larger vials, 16mL
  • 12 places vials holder (small and medium)
  • 6 places vials holder (large)
  • 7 places vials holder (4 small, 3 large)
  • Analytical balance
  • NCAL, calibration kit
  • Keyboard and mouse
  • Printer
Security
  • Emergency stop button
  • Safety thermocouple (separate from the oven’s thermocouple)
  • Audible alarm
  • Safety shutdown point: 600°C (configurable)
  • Shutdown on failure or out-of-tolerance nitrogen flow
  • Automatic shutdown beyond 300 mins
Software
  • 7’’ Capacitive touchscreen
  • Possibility to connect external mouse and keyboard
  • Windows 11
  • Multilingual (FR/EN)
  • Data export via USB, e-mail, RS-232, or print on a USB-connected PC printerAuto-diagnostic
  • Self-diagnostic
  • User-friendly software interface
  • 3 access levels (including operator profile)
  • Temperature, heating power and nitrogen flow graphs
  • > 1 000 results & > 1 000 product files
  • Deferred weighing: once amples moved to desiccator; new run can start, weights entered afterwards.
Site requirements
  • Power supply: 230V (±5%) – 50/60Hz, 1 600 W (7.5A)
  • Dimensions (mm) : 605(W) x 345(D) x 445(H)
  • Weight: 37kg
  • Nitrogen: 9 bar (900kPa) max upstream, 2 bar (200kPa) in service

ASTM D4530 Method: Analysis of carbon residues by pyrolysis

Related products
Move up a gear
No longer leave the accuracy of your analyses to chance.
Contact us today and take your laboratory to the next level.
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.